beautypg.com

Connecting the system to power, Calculating the system power requirement, Caution – Measurement Computing WaveBook rev.5.3 User Manual

Page 48

background image

Connecting the System to Power

Calculating the System Power Requirement

CAUTION

An incorrect use of power can damage equipment or degrade performance. Prior to

connecting your devices to power, calculate your system’s power requirements.

CAUTION

Do not daisy-chain the power connections of more than three WBK10A modules.

Daisy-chaining a power connection to a fourth module will exceed the power

connector’s 5 amp current limit.

It is important to supply your system with adequate, reliable power. For this reason, you need to know

your system’s power requirement. Computing power use is also important when using batteries to power

modules, as you will need to know a safe runtime before recharging is required.

The following statements relate to system power. They should be reviewed before

proceeding.

Higher voltages draw fewer Amps for the same power.

Remember: Watts = voltage x current (W = V*I).

The TR-40U power adapter provides power that is sufficient for the WaveBooks

and WaveBook modules. You do not need to make power requirement

calculations unless you intend daisy-chaining units, or you have a critical battery

runtime.

Do not overload your power supplies. TR-40U power adapters are limited to

2.2 amps. However, you can use more than one TR-40U, as indicated in one of

the upcoming daisy-chain examples.

Current drawn from other sources, such as car batteries, can be estimated from

the following WaveBook Product Current Requirements table.

Use the current requirements and worksheet tables to calculate your system’s total power requirement.

Take the appropriate amperage values from the first table to fill in the second table; then perform the

indicated multiplication and addition operations to calculate the amperage for all units in your system.

2-28 System Setup and Power Options

979194

WaveBook User’s Manual