beautypg.com

13 - specifications, General specifications, Analog specifications – Measurement Computing ZonicBook 618E rev.3.4 User Manual

Page 97

background image

ZonicBook/618E

918895

Specifications 13-1

Specifications - ZonicBook/618E

13

Specifications are subject to change without notice.

General Specifications

Environment:

Operating: 0˚ to 50˚C, 0˚ to 95% RH, non-condensing
Storage: -20˚ to 70˚C

Power Consumption: 2.3A max @ 15 VDC
Input Power Range:
10 to 30 VDC
Vibration: MIL STD 810E, Categories 1 and 10
PC Communication: 10/100BaseT Ethernet
Dimensions:
285 mm W x 220 mm D x 70 mm H; (11” x 8.5” x 2.70”)
Weight: 2.38 kg (5.25 lbs.)
Warm-up: 30 minutes to rated specifications
Throughput rate
: >2 Mbytes/s
Internal Data Storage:
1 Msample built-in, 64 Msample optional (factory installed)
SYNC Input:
Reserved for future use.


Analog Specifications

Analog Inputs:

Channels: 8 single-ended input channels, expandable up to 56 channels via six WBK18 modules
Input Connector: 1 BNC per channel
Input Impedance: 200k Ohm (single-ended)
Input Coupling:
AC or DC; software programmable per channel
High-Pass Filter:
0.1 Hz or 1 Hz; software programmable per channel
Input Ranges:

±25V (DC coupled only), ±5V, ±2.5V, ±1V, ±500mV, ±250mV, ±100mV, ±50mV, ±25mV;

software programmable per channel

Overrange Indication: Front panel LEDs, one per channel, software status

Low-Pass Filter (software programmable per channel)

Type: 8-pole Butterworth with simultaneous sample-and-hold (SSH)
Cutoff Frequency (Fc): 10 Hz to 200 kHz in 1-2-5 progression
Alias Rejection:
75 dB min
Channel-to-Channel Phase Matching*: 1

° typ, 2° max

Unit-to-Unit Phase Matching*: 1

° typ, 2° max

SSH latency: 100ns max

Amplitude Accuracy**:

±0.5 dB (F

in

≤ F

c

/2)

Total Harmonic Distortion: -70 dB typ
ICP Bias Source: 4mA, 24V compliance (on/off software programmable per channel)
ICP Fault Detection Thresholds: <1V (short), >25V (open)
ICP Fault Indication: Front panel LED per channel, software status
Coupling: AC, AC with ICP; or DC. Programmable on a per-channel basis

* Conditions for Low-Pass Filter Phase-Matching

8-pole LPF mode, 0.1 Hz or DC HPF mode
1 Hz

≤ F

in

≤ F

c

/2

200 Hz

≤ F

c

≤ 20 kHz

** Condition for Amplitude Accuracy

For Fin < 20 kHz

This manual is related to the following products: