Measurement Computing ZonicBook 618E rev.3.4 User Manual
Page 90
Current Requirements (in Amps)
Products
DBK30A
14 VDC
DBK30A
28 VDC
DBK34A
12 VDC
DBK34A
24 VDC
TR-40U
15 VDC
ZonicBook/618E (alone)
2.5 1.3 2.8 1.4 2.3
WBK18
1.4 0.75 1.6 0.80 1.3
WBK30
0.01 0.005 0.01 0.006 0.01
Worksheet for Power Requirements
Product
Qty
×
Amps
=
Totals
WaveBook/516E
WBK18
× =
WBK30
× =
Maximum
Amps
ZonicBook/618E Product Power Supplies
Item
Name/Description
Capacity
TR-40U
AC Power Adapter (shipped with ZonicBook/618E &
WBK18)
90-264 VAC input;
3.33 A @ 15 VDC
DBK30A
Rechargeable Battery/Excitation Module (optional)
12-14 VDC, or 24-28
VDC
3.4 A-hr @ 14 VDC
DBK34A
UPS (uninterruptible Power Supply)/Battery Module
(optional)
12 VDC, or 24 VDC
5.0 A-hr @ 12 VDC
Other
10 to 30 VCD source, such as a vehicle battery via a
CA-116 cable.
Depends on source
Input voltage to the ZonicBook/618E and to the WBK18 expansion modules must be in the range of
10 to 30 VDC and can come from an AC-to-DC adapter or from another source, such as a battery.
Available AC-to-DC adapters include the TR-40U, which has an input of 90 to 264 VAC and an output
of 3.33 amps @ 15 VDC.
Battery options include the DBK30A, DBK34A, and other 10 to 30 VDC sources, such as car batteries.
The DBK30A provides 14 VDC and when fully-charged has a storage capacity of 3.4 A
⋅hr; car batteries
have much higher capacities. The basic formula for battery life is:
Runtime (hr) = Battery capacity (A⋅hr) / Current load (A)
Battery life and performance depend on various factors including battery type, condition,
charge level, and ambient temperature. Be sure you consider these factors, especially
when runtime is a critical.
Reference Note:
For information regarding alternate power supplies, refer to Appendix C, Power Options.
11-6 System Expansion
929795