beautypg.com

8 - trigger input & clock input, Trigger input, Clock input – Measurement Computing ZonicBook 618E rev.3.4 User Manual

Page 75: Trigger input & clock input 8

background image

Trigger Input & Clock Input

8

Trigger Input – Available for eZ-Analyst only

Clock Input – Reserved for future use

Trigger Input and Clock Input BNC Connectors

The top row of the front panel includes a Trigger Input BNC and Clock Input BNC connector. Other connectors
are discussed elsewhere.

Trigger Input

The ZonicBook/618E includes a Trigger Input BNC. This TTL compatible trigger input applies only to the
ZonicBook/618E when eZ-Analyst is being used.

The Trigger Input can be used to:

• trigger the start of a data acquisition
• trigger the start of recording data
• trigger both the start of an acquisition and the start of data recording at the same instant

Connector: BNC, labeled TRIGGER INPUT

Input Signal Range: 0 to 5V, TTL compatible

Input Characteristics: TTL compatible with 10K Ohm pullup resistor

Latency: 300 ns max

Reference Notes:

¾

For information regarding the use of an analog input channel (CH1 through CH8)

as a trigger, refer to the Analog Triggers section of chapter 6 in this document.

¾

For information regarding eZ-Analyst and trigger input, refer to the eZ-Analyst

Software Reference Manual, p/n 1086-0922.

Clock Input

The Clock Input BNC is reserved for future use.

ZonicBook/618E

929895

Trigger Input & Clock Input 8-1

This manual is related to the following products: