Emissions/Immunity Conditions

The specific immunity conditions for CE compliance vary by product; but general immunity conditions
include:

• Cables must be shielded, braid-type with metal-shelled connectors. Input terminal connections are to

be made with shielded wire. The shield should be connected to the chassis ground with the hardware
provided.

• The host computer must be properly grounded.
• In low-level analog applications some inaccuracy is to be expected when I/O leads are exposed to

RF fields or transients over 3 or 10 V/m as noted on the Declaration of Conformity.

Using Shielded BNC Connectors for CE Compliance

Certain Declarations of Conformity identify specific cables and connectors that must be used to meet CE
requirements. CE compliant BNC-equipped cards and modules have BNC connectors that are insulated
from high voltage sources, including electrostatic discharges (ESD). Such voltages could enter the
circuitry through the exposed conductive surface of a connector, possibly resulting in damage to
components.

O-Ring

Dust Cap

Shielded
BNC Connector

Shielded BNC Connector (with O-Ring) and PVC Dust Cap

To meet CE requirements, PVC dust caps (p/n CN-96) must cover all unused BNC connectors. When dust
caps are not in place, special coaxial cables (with insulated end-connectors and rubber O-rings) must be
used. Note that part number 418-0800 includes two cables (with shielded BNC connectors at each end),
and four insulating O-rings.

Properly installed connectors and dust caps ensure the metallic surfaces of the connectors are not exposed
to undesirable electrical charges.

CE Compliance for ZonicBook/618E System Expansion

ZonicBook/618E units are CE Compliant at the time they leave the factory. They will remain in
compliance providing that the conditions stated on their associated Declaration of Conformity (p/n 1109-
0740) continue to be met.

As this document goes to press ZonicBook/618E units are approved for system expansion with up to six
WBK18 modules. Prior to expanding the system, refer to the Declaration of Conformity for the WBK18
module. Be aware that connecting a non-CE Compliant module or card to your system will make the
system non-compliant.

A few general rules of thumb regarding the expansion of CE compliant ZonicBook/618E systems are:





Use short cables, e.g. CA-35-12, CA-202.

When possible use shielded cables.
When assembling or disassembling components, take ESD precautions,
including the use of grounded wrist straps.
Ensure that the host computer is CE Compliant.
Review the most recent Declaration of Conformity for each device.
Ensure all system components are properly grounded.