6-6 Analog Signals

878595

ZonicBook/618E

IN (kHz)

F

IN

kHz

F

IN

(kHz), Zoom-In to Filter Cutoff Frequency Region

F

IN

kHz

Filter settings are discussed in the separate software documentation.

DC Accuracy

The bypass filter scenario is most commonly used in applications where DC signal information is
important.

DC Accuracy, Excluding Noise

Applies to 2-Pole and Bypass Filter Modes

Accuracy at 0° to 50°C (32° to 122°F)

± % of Reading

Range

Typical

Maximum

Offset

±25V

0.5 1.0

±15mV

±5V

0.15 0.3

±3mV

±2.5V

0.15 0.3

±2mV

±1V

0.15 0.3

±1mV

±500mV

0.15 0.3

±1mV

±250mV

0.15 0.3

±1mV

±100mV

0.15 0.4

±1mV

±50mV

0.15 0.4

±1mV

±25mV

0.2 0.4

±1mV

These numbers are valid for 1 year after calibration and over the
entire operating temperature range of the unit.