Measurement Computing ZonicBook 618E rev.3.4 User Manual
Page 54

-4.5
-4
-3.5
-3
-2.5
-2
-1.5
-1
-0.5
0
0
25
50
75
100
125
150
175
200
225
250
F
)
IN (kHz
G
ain (
d
B
)
-0.3
-0.25
-0.2
-0.15
-0.1
-0.05
0
0
5
10
15
20
25
30
35
40
45
50
F
G
ain
(
d
B
)
6-6 Analog Signals
878595
ZonicBook/618E
IN (kHz)
F
IN
kHz
F
IN
(kHz), Zoom-In to Filter Cutoff Frequency Region
F
IN
kHz
Filter settings are discussed in the separate software documentation.
DC Accuracy
The bypass filter scenario is most commonly used in applications where DC signal information is
important.
DC Accuracy, Excluding Noise
Applies to 2-Pole and Bypass Filter Modes
Accuracy at 0° to 50°C (32° to 122°F)
± % of Reading
Range
Typical
Maximum
Offset
±25V
0.5 1.0
±15mV
±5V
0.15 0.3
±3mV
±2.5V
0.15 0.3
±2mV
±1V
0.15 0.3
±1mV
±500mV
0.15 0.3
±1mV
±250mV
0.15 0.3
±1mV
±100mV
0.15 0.4
±1mV
±50mV
0.15 0.4
±1mV
±25mV
0.2 0.4
±1mV
These numbers are valid for 1 year after calibration and over the
entire operating temperature range of the unit.