AMETEK MX CTSL User Manual

Page 79

background image

User Manual

MX-CTSL Compliance Test System

California Instruments

Revision J

79

5.9

Printing Results

The Report button allows you to print both on-screen graphs displays and current and voltage harmonics
test result data to an MS Word document. This provides a test report that displays the test setup, the Pass
or Fail result and a graph of the voltage and current waveform as well as the worst case values for each of
the current harmonics. The Report button is available at the end of a test run or any time a test is aborted
with the Stop button.

Information included on the harmonic test report will be a function of the selected test mode and EUT
connection stage. The following fields provide this information:

Heading:

Test category:

Selected Stage and EUT type.

Current Test Result Summary

Test Result

Overall Pass or Fail result

Highest parameters recorded during the test.

V_RMS

EUT Voltage Line to Neutral

I_Peak

Maximum peak current recorded

I_Fund

Maximum fundamental current recorded (I1)

Power

Maximum power level recorded

I_RMS

Maximum RMS current recorded

Crest Factor

Power Factor

Harmonic Currents Table:

Harmonic number (#)

Only harmonics for which limits apply are shown.

Harms (Avg)

Average harmonics value (1.5 sec smoothing
filter applied)

100% Limit

Absolute current value of 100 % limit for each
harmonic

% of Limit

Percent of 100 % limit for each harmonic

Harms (Max)

Maximum harmonic value recorded during test.

150% Limit

Absolute current value of 150 % limit for each
harmonic

% of Limit

Percent of 150 % limit for each harmonic
maximum

Status

Individual harmonics pass or fail result (based on
100% and 150% limits)

Voltage Source Verification Data

The third page of the report contains data on the source voltage. If needed, the CTSMXL software will
compensate for voltage distortion that would affect the harmonic current results. This will be indicated.

Note that the overall EUT test results is repeated in the header of this page. This pass or fail does not
apply to the AC source.

This manual is related to the following products: