13.

EN 61000-4-29p DC Dips and Interruptions Immunity Test

(Pre-compliance)

Figure 13-1: EN 61000-4-29 Test Window

The EN 61000-4-29 implementation in the MXGUI is based on the EN 61000-4-29 standard dated march
24, 2000. This test is offered as a pre-compliance level test only. Not all the DC supply rise and fall
time and maximum current requirements can be met under all conditions. (See paragraph 13.8 for details.)
Test parameters can be changed if needed using the parameter data entry grid. Several generic test
parameter files are distributed with the MXGUI program. These files can be edited as well. EN 61000-4-29
Test setup parameter files use a “.429” extension. More than one .429 file may reside in the IEC_Test
directory or any other directory. The operator may select the one to run from the File, Open dialog.