beautypg.com

AMETEK MX CTSL User Manual

Page 106

background image

User Manual

California Instruments

9.3

Test Sequence

The test sequence implemented
commonly found phenomena
product for immunity from such fluctuations.

Voltage fluctuations are applied at different levels
the product class and select the
specified by the selected test

The test sequence can be programmed
change from the nominal voltage in %, the number of
periods) and the delay between
repeated 1 or more times. (A Delay of
data setup are shown in the figure
data entry grid:

Unom (%)

Test Level

100

Note that the sequence of voltage
100 % Unom, the voltage changes
level (%) is set to a negative value

These parameters can be entered
test class, select the File, Open menu.

C:\Program Files\California Instruments

The following four EN 61000-4

File Name

Generic IEC 1000-4-14 Test Class

Generic IEC 1000-4-14 Test Class

The user may create as many
using the File, Save As… menu entry.

MX-CTSL

Revision H

implemented by this test consists of a series of voltage fluctuations

on the public utility power grid. The objective of this test is to evaluate

product for immunity from such fluctuations.

Voltage fluctuations are applied at different levels for different product categories.

the appropriate test level. During the test run, voltage

mode and for specified periods of time.

programmed by setting the nominal voltage in %, the test level which is the

change from the nominal voltage in %, the number of times each test level repeats

between each set of test levels. Finally, each step in the

or more times. (A Delay of 1 means the set it only run once.) The various

figure below. In this example, the following parameters

Level (%)

Level Repeat

Delay

8

3

60

Figure 9-2: EN 61000-4-14 Test Sequence

voltage changes is different if the Unom is at 100 %

changes occur in both positive and negative directions (alternating).

value and Unom is 100 %, the voltage changes are

entered on screen by the operator or loaded from disk.

pen menu. The default location for IEC test files is:

Instruments\Mxgui\IEC_Test

4-14 test files are distributed with the MXGUI program:

Class 2.414_Fluc

Class 3.414_Fluc

many test files as desired using the data entry grid. Changes

s… menu entry.

CTSL Compliance Test System

103

fluctuations consistent with

objective of this test is to evaluate a

categories. The user must determine

voltage changes are applied as

nominal voltage in %, the test level which is the

repeats (consecutive 5 second

the sequence data grid can be

various columns in the test

parameters were specified in the

Repeat

1

than at any other value. At

negative directions (alternating). If the test

are reversed.

disk. To load a new test or

The default location for IEC test files is:

program:

Test Class

Class 2

Class 3

Changes made can be saved

This manual is related to the following products: