AMETEK MX CTSL User Manual
Page 139

User Manual
MX-CTSL Compliance Test System
Line Input
Line Voltage
115 Vac±10% / 230 Vac±10%
Line Current
200 mA @ 115 Vac / 60 Hz
110 mA @ 230 Vac / 50 Hz
Line Frequency
47-63 Hz
Fuse Rating
0.25 A slow acting @ 115 Vac
0.125 A slow acting @ 230 Vac
Mechanical
Dimensions
Including base HxWxD (inch)
5.25” x 19” x 22”
HxWxD (mm).
133 x 480 x 560
Weight
Lbs / kg
55 lbs / 25 Kg
Chassis Slides
General Devices C300S-118-U/B308 BK
15.8
OMNI-3-37MX Environmental
Parameter
Specification
Operating Temp:
0 degrees to +40 degrees Celsius.
Protection:
OMNI will dis-engage if internal heat sink temperature
exceeds 95° C. Re-engages once temperatures
declines below threshold.
Storage Temp:
0 degrees to +70 degrees Celsius.
Humidity:
Operating:.
≤
90% RH up to 40
°
C
Storage:
≤
90% RH up to 40
°
C,
≤
75% RH up to 70
°
C.
Creepage and Clearance:
Rated for Pollution Degree 2.
Insulation:
Rated to Installation Category (Over voltage Category)
II
Vibration:
Designed to meet NSTA 1A transportation levels.
Shock:
Designed to meet NSTA 1A transportation levels.
15.9
OMNI-3-37MX Regulatory
Parameter
Specification
Electromagnetic Emissions and
Immunity:
Designed to meet EN50081-1 and EN50082-1
European Emissions and Immunity standards as
required for the “CE” mark.
Acoustic Noise:
65 dBA maximum. (@ 1 meter)
Safety:
Designed to meet EN61010-1 European safety
standards as required for the “CE” mark.
California Instruments
Revision H
136