B. Criteria
Task
Description
PoE-Overload
Icut: Configure the overload current detection
range.
Tovld: Configure the overload time limit in
milliseconds.
Power Cut Time: The time range to cut the power.
Power meter Setting: this function is available only when XM-2WL1 modules are installed. The
function is to test the electric parameters of DUT. Please select the XM-2WL1 module from the
scroll-down menu.
Power Setting: This function allows you to check DUT‟s power status during the test. Please note
that if there is no power testing module card installed on the chassis, the Power Setting section will
be gray and thus unable to access.
Enable Power Monitor & Check: Click this check-box to enable Power Setting function. You
can set which port on the power testing module card to use by clicking the scroll-down menu.
RMS Voltage/Current, Peak Voltage/Current, and Active Power: You can set the criteria for
power testing in these fields.
C. Help
You can see the system status of your task on
this field.