3.3 Setup Software Operating Procedure
34
4. When you select Inspect check-digit, the following information appears:
The above setting sends the data together with the check-digit.
Modulus 43 is used to calculate the check-digit.
5. When you select Max code length output, the following information appears:
Set Effective and Starting from 1 to 32.
6. When you select Specify label orientation, the following information appears:
7. Click the
Return
button or press
[ENTER]
to return to the main setting
screen.
Max. code length output --> Page 23
Specify label orientation --> Page 16
3.3.4
ITF setup
1. Setup is the same as the CODE39 setup. However, ITF does not have Start/
stop character.
2. Set code length using a even value within the range from 2 to 32.
3. Modulus 10/Weight3 is used to calculate the check-digit.
When reading the standard distribution code (bar code on the corrugated fiber-
board box), set the code length to 14 or 16 and select the Inspect check-digit.
3.3.5
Setup for Industrial 2 of 5 and COOP 2 of 5
1. Setup is the same as the CODE39 setup. However, a start/stop character
and Inspect check-digit are not provided.
2. Set code length within the range from 1 to 32.