Hardy HI 2151/30WC Single-Scale Controller User Manual
Page 113
Chapter 7 - Operation
7-15
Dip Switch #1 (S3)
Position Bit
Function ("ON" position)
1
80
Multi-drop
2
01
Calibrate toggle (Non NBS applications)
3
40
Serial command lockout (Requests OK)
4
02
Option menu lockout
5
04
Set point menu lockout
6
20
Spare
7
08
Lb/Kg, N/G, Tare and Zero lockout
8
10
Zero track
SUB-CMD
Meaning
Dip Switch #2 (S2)
Position Bit
Function ("ON" position)
1
08
Ignore incoming checksums
2
10
On is averaged peak hold, Off is instantaneous
peak
hold
3
20
NBS
4
04
NBS Re-Cal toggle switch
5
40
Spare
6
80
Spare
7
02
Spare
8
01
Spare
SUB-CMD
Meaning
E
Excitation Monitor
ERR
OKK
OFF
3. E IT
®
serial command
Results
Description
OKK
No Errors Detected
ERR
Error Detected
OFF
IT is Not Turned On
4. I
Load cell input command. Used primarily to
check the analog input section of the unit. The command
will return the number of averaged counts that is currently
being read from the load cell input.