Recorder Pen Moves in Wrong Direction

Output polarity reversed

• Reverse leads

Viscometer Wil Not Communicate with PC

✓ Check the comm port and make sure the correct port is being utilized.
✓ Check the interconnecting cable for proper installation

✓ Check the Options menu and make sure the PC PROG is set to either “ON” or “OFF” in

accordance with the operating instructions for the program/procedure in use.

If the above do not rectify the problem, do the following:

✓ Shut off viscometer

✓ Attach printer.

✓ Press and hold the

MOTOR ON/OFF

and

ENTER/AUTORANGE

keys simulanteously while turning

the viscometer power on. Figure K-1 appears on the viscometer display.

VISCOMETER SETUP
ENTER TO START

Figure K-1

✓ Press

ENTER/AUTORANGE

key and Figure J-2 appears.

PRINT PARAM RAM?
NO#$ THEN ENTER

PRINT PARAM RAM?
YES#$ THEN ENTER

"

t

Figure K-2

✓ Scroll to YES and press the

ENTER/AUTORANGE

key. Figure K-3 appears.

READY PRINTER
ENTER TO START

Figure K-3

✓ Press ENTER/AUTORANGE. Information similar to Figure K-4 will appear on your printer

output.