beautypg.com

Brookfield DV2+Pro Viscometer User Manual

Page 15

background image

Brookfield Engineering Labs., Inc.

Page 15

Manual No. M03-165-F0612

Note: The DV-II+

Pro

will remember the Spindle Entry Code which was in use

when the power was turned off.

Pressing the SELECT SPINDLE key will temporarily display the current selected spindle code

in place of temperature and cause the character S to begin to blink . It will blink for about three

seconds. If the UP or DOWN ARROW keys are pressed (while S is blinking), the two character

spindle value to the right of the S character will begin to change (in either an increasing or

decreasing direction depending upon which ARROW key is pressed) for each press of the key.

If the ARROW key is pressed and held, the display will scroll through the spindle codes for as

long as the ARROW key is depressed. When it reaches the last item in the list (either at the top

or bottom of the list) the spindle code displayed will “roll-over” to either the first or last spindle

code and the scroll action will continue.

When the desired spindle code is displayed, release the ARROW key to halt further scrolling.

Press the SELECT SPINDLE key once again. This will cause the S character to cease blinking

and the new spindle code will be accepted for use in viscometer calculations. After 3 seconds

the current spindle code will be replaced by the temperature display.

Note: You have approximately three seconds in which to press the SELECT

SPINDLE key before the blinking stops. If you fail to press the SELECT

SPINDLE key before the blinking stops you will have to repeat the above

steps and re-select the desired spindle.

The DV-II

+

Pro will begin to calculate using the new spindle parameters as soon as the

SELECT

SPINDLE

key is pressed the second time.

Note: The number 99 spindle is for use with special spindles when using

Brookfield’s Rheocalc computer program. Refer to the Rheocalc operator

manual for further information on using “99” spindles.

The DV-II+Pro may also be programmed at Brookfield Engineering for “special” user spindles.

These “special” spindles will appear on the spindle scroll list starting with designation “AA”

and continuing through “AZ”. Contact Brookfield Engineering regarding your needs for special

spindles.

II.3

SET

SPEED

Speed Selection, Setting, Running

There are 54 speeds programmed into the DV-II+Pro. These speeds correspond to the standard

LVT, RVT, HAT and HBT dial models (18 possible speeds altogether) plus 36 additional speeds.

The DV-II+Pro comes with the Sequential Speed Set already selected (see Appendix G). The

speed set will start at speed 0.0. It will then scroll up through the LV speeds, pass through speed

0.0 again, and then scroll up through the RV speeds, pass through speed 0.0 again and then repeat

the above sequence.

The DV-II+Pro can also be configured by the operator to interleave the LV and RV speeds. See

Section IV.2.3 on Setup for a description of how to install the Interleave Speed Set.

A complete list of speed sets and custom speeds is included in Appendix G. The DV-II+Pro can be

programmed to select up to 19 of the 54 speeds for use at any one time. Speed 0.0 is automatically

included as one of the nineteen (19) speeds. See Section IV. 2.3.2 on Setup for a description of