Brookfield DV2+Pro Viscometer User Manual
Page 31

Brookfield Engineering Labs., Inc.
Page 31
Manual No. M03-165-F0612
AUTORANGE
key to select the flashing option.
É
DCGS UNITS
DSEQUENTIALê
1
DCGS UNITS
DINTERLEAVE
DCGS UNITS
DCUSTOM SPEEDSê
1
Figure IV-9
The speeds available in each of the above options are listed in Appendix G. The DV-II+Pro is
initially set up with the Sequential Speed Set at Brookfield prior to shipment.
IV.2.3.1 LV/RV Speeds
In the case of Sequential or Interleave, a press of the
ENTER/AUTORANGE
key immediately selects
that option and exits the SETUP option menu.
IV.2.3.2 Custom Speeds
Pressing the
ENTER/AUTORANGE
key when Custom Speeds is displayed results in the following
screen display:
.01
h
°0.0
i
Figure IV-10
Fifty-four (54) speeds are available for custom speed set selection (see Appendix G). You are
allowed to select up to nineteen (19) of these (54) available speeds. Speed 0.0 is automatically
included as one of the nineteen (19) speeds. Selecting a speed (or deleting a speed) is accomplished
by pressing the
SET SPEED
key while the desired speed is blinking. This will cause an asterisk
to appear (or to disappear if the speed is being cleared) to the left of the speed. Trying to select
more than nineteen (19) speeds will result in a “beep” for each press of the
SET SPEED
key when
over this limit. You may scroll up or down through the speed set in selecting speeds. When done,
a press of the
ENTER/AUTORANGE
key will take you back to the default screen with the asterisked
speeds now comprising the custom speed set. Regardless of order chosen, speeds will appear in
ascending order for run selection, beginning with speed 0.0 RPM.
IV.2.4 Printer Output Port
Scrolling to the printer port option presents the following:
É
DSEQUENTIAL
DPRINT SERIAL
DSEQUENTIAL
DPRINT PARALLEL
"
Figure IV-11
#
$
#
$
#
$
#
$
#
$