A.2 Setup

1. Be sure that the Viscometer is securely mounted

to the Laboratory Stand, leveled and zeroed

with no cone or cup attached and 0% torque is

displayed.

2. Figure A-2 shows a typical water bath setup.

Connect the sample cup inlet/outlet ports to the

water bath inlet and outlet and set the bath to

the desired test temperature. Allow sufficient

time for the bath to reach the test temperature.

3. The Viscometer has been supplied with a

special cone spindle(s) which contains the

Electronic Gap Setting feature. The “CPE” or

“CPA” part number designation on the cone

spindle verifies the Electronic Gap Setting

feature.

Note:

The “CPE” or “CPA” cone or cup

cannot be used with earlier DV-II+

cone/plate Viscometers (below S/

N50969) which do not have the

electronic gap setting feature.

4. With the motor off, thread the cone spindle by

using the spindle wrench to secure the viscom-

eter coupling nut (see Figure A-3); gently push

up on the coupling nut and hold this securely

with the wrench. Thread the cone spindle by

hand.

Note:

Left Hand Threads.

5. Attach the cup, taking care not to hit the cone

with the cup (Figure A-4), by positioning the

cup against the mic ring and swinging the ten-

sion bar under the cup. The tension bar should

have plastic tubing in place.

Bath/Circulator

Bath

Inlet

Bath

Outlet

Sample

Cup

Cup

Outlet

Cup

Inlet

Figure A-3

Spindle

Wrench

Cone Spindle

These surfaces

must be clean!

Coupling Nut

Micrometer

Adjustment

Ring

Do Not hit the

CONE with the CUP!

Tension Bar

Figure A-2

Figure A-4