Manual No. M03-165-F0612

Viscometer Will Not Return to Zero

Viscometer is not level

• Check with spindle out of the sample

• Adjust the laboratory stand

Pivot point or jewel bearing faulty

• Perform an Oscillation Check*

Remove the spindle and turn the motor OFF; select display to % torque mode.

Gently push up on the viscometer coupling.

Turn the coupling until the digital display reads 10-15 on the % display.

Gently let go of the coupling.

Watch the digital display; you should see a “run” of numbers next to the %; the “run” of

numbers should ultimately stop at 0.0 (+/- 0.1).

If the digital display does not return to ZERO, the unit most likely is in need of service.

• Perform calibration check (See Appendix E)

• Contact Brookfield Engineering Laboratories, Inc. or your Brookfield dealer for repair (See

Appendix N).

* This procedure does not apply to instruments with ball bearing suspension. (See section I.3)

Display Reading Will Not Stabilize

Special characteristic of sample fluid. There is no problem with the viscometer.

• Refer to Appendix C

Check for erratic spindle rotation.

• Verify power supply

• Contact Brookfield Engineering Laboratories, Inc. or your Brookfield dealer for repair.

Bent spindle or spindle coupling.

• Contact Brookfield Engineering Laboratories, Inc. or your Brookfield dealer for repair.

Temperature fluctuation in sample fluid.

• Use temperature bath for control

No Recorder Response

Be Sure the viscometer is not at ZERO reading.

Be sure the recorder is ON and not on STANDBY.

Verify the range settings.

Check cable leads for clean connection.

Verify cable connections (see Appendix H).