To stop continuous printing, press the PRINT key for one (1) second. The flashing P will

disappear on the viscometer display.

Figure II-21 is an example of the print strings for CGS and SI units.

For the case of CGS units with non-exponential results:

and CGS units with exponential results.

Similarly, for SI units with non-exponential results.

and SI units with exponential results.

M = Torque T = Temperature Z = Time

1

10

20

30

40

50

60

70

80

RPM=XXX M=XXXXX S=XX %=XXX.X cP=XXXXX D/CM2=XXXXX 1/SEC=XXXXX T=XX.XC ZXX=XX

RPM=XXX M=XXXXX S=XX %=XXX.X cP=XXXeX D/CM2=XXXeX 1/SEC=XXXXX T=XX.XC ZXX=XX

1

10

20

30

40

50

60

70

80

RPM=XXX M=XXXXX S=XX %=XXX.X mPas=XXXXX N/M2=XXXXX 1/SEC=XXXXX T=XX.XC ZXX=XX

1

10

20

30

40

50

60

70

80

1

10

20

30

40

50

60

70

80

RPM=XXX M=XXXXX S=XX %=XXX.X mPas=XXXeX N/M2=XXXeX 1/SEC=XXXXX T=XX.XC ZXX=XX

Figure II-21

When printing via the parallel port, please note that if a printer is not attached to the viscometer,

the following screen appears:

PRINTER ERROR

CHECK CONNECTION

Figure II-22

II.9 External Control Mode

The DV-II+Pro Viscometer can be used in conjunction with Brookfield software, Rheocalc (V2.4

or higher). Through Rheocalc, all viscometer functions are controlled by the computer. The DV-

II+Pro must be set to the external control mode to allow for proper communication with Rheocalc.

To configure the external control mode, connect cable DVP-80 to the serial port on the DV-II+Pro

before turning on the DV-II+Pro. With the DVP-80 cable in place, the DV-II+Pro will present the

screen shown in Figure II-23 when it is turned on. If external control is selected, the DV-II+Pro

will display Figure II-24 and only accept control commands from Rheocalc software.

hEXTERNAL MODE
iSTANDALONE MODE

Figure II-23

V6.3

RV

EXTERNAL MODE

Figure II-24