The shear rate of a given measurement is determined by: the rotational speed of the spindle, the size

and shape of the spindle, the size and shape of the container used and therefore the distance between

the container wall and the spindle surface.

A repeatable viscosity test should control or specify the following:

1)

Test temperature

2)

Sample container size (or spindle/chamber geometry)

3)

Sample volume

4)

Viscometer model

5)

Spindle used

6)

Whether or not to attach the guard leg

7)

Test speed or speeds (or the shear rate)

8)

Length of time or number of spindle revolutions to record viscosity.

9)