Figure f-1 - brookfield guard leg, Figure f-2 - brookfield ez-lock guard leg – Brookfield DV2+Pro Viscometer User Manual
Page 87

Brookfield Engineering Labs., Inc.
Page 87
Manual No. M03-165-F0612
should be known that this type of torque reading will not convert into a correct centipoise value when
using a Brookfield factor if the boundary conditions are not those specified by Brookfield.
The guard leg is a part of the calibration check of the Brookfield LV and RV series Viscometer/Rheometer.
Our customers should be aware of its existence, its purpose and the effect that it may have on data.
With this knowledge, the viscometer user may make modifications to the recommended method of
operation to suit their needs.
4 11/16
1 7/16
3 3/16
5 9/32
Figure F-1 - Brookfield Guard Leg
B-21KY
Guard Leg
B-20KY
Guard Leg
5 7/8
5 5/32
For RV Torque
For LV Torque
3 3/16
1 7/16
Figure F-2 - Brookfield EZ-Lock Guard Leg