II.6 Out of Range

The DV-II

+

Pro gives indications for out-of-range operation. When % (Torque) readings exceed

100% (over-range), the display changes to that shown in Figure II-15; EEEE will also appear in

the display for shear stress:

cP EEEE

20.1C

10 RPM

% EEEE

Figure II-15

You must change either speed or spindle to correct this condition. If you operate at spindle speeds

that produce % (Torque) below 10.0 %, the DV-II+Pro flashes the % (Torque), cP (Viscosity), SS

(Shear Stress) and SR (Shear Rate) as shown in Figure II-16:

cP 12.4

20.1C

10 RPM

% 8.2

Figure II-16

Negative % (Torque) will be displayed as shown in Figure II-17:

cP ----

20.1C

10 RPM

% -2.2

Figure II-17

Figure II-18 is an example of the printed output of each of the above conditions.

Normal Operation:

RPM=50 M=RV S=29 %=51.4 cP=10280

D/CM2=1285

1/SEC=12.3 T=20.1C Z00:30

Over-Range Operation (>100% torque) (see Fig. 15):
RPM=50 M=RV S=29 %=EEEE cP=EEEE

D/CM2=EEEE

1/SEC=12.3 T=20.1C Z00:30

Under-Range Operation (<10% torque) (see Fig. 16):
? RPM=50 M=RV S=29 %=5.2 cP=1040

D/CM2=130

1/SEC=12.3 T=20.1C Z00:30

Negative Torque Operation (see Fig. 17):
RPM=50

M=RV S=29 %=-0.1 cP=---- D/CM2=----

1/SEC=12.3 T=20.1C Z00:30

M = Torque Range

T = Temperature

Z = Time

Figure II-18