II.6 Out of Range
The DV-II
+
Pro gives indications for out-of-range operation. When % (Torque) readings exceed
100% (over-range), the display changes to that shown in Figure II-15; EEEE will also appear in
the display for shear stress:
cP EEEE
20.1C
10 RPM
% EEEE
Figure II-15
You must change either speed or spindle to correct this condition. If you operate at spindle speeds
that produce % (Torque) below 10.0 %, the DV-II+Pro flashes the % (Torque), cP (Viscosity), SS
(Shear Stress) and SR (Shear Rate) as shown in Figure II-16:
cP 12.4
20.1C
10 RPM
% 8.2
Figure II-16
Negative % (Torque) will be displayed as shown in Figure II-17:
cP ----
20.1C
10 RPM
% -2.2
Figure II-17
Figure II-18 is an example of the printed output of each of the above conditions.
Normal Operation:
RPM=50 M=RV S=29 %=51.4 cP=10280
D/CM2=1285
1/SEC=12.3 T=20.1C Z00:30
Over-Range Operation (>100% torque) (see Fig. 15):
RPM=50 M=RV S=29 %=EEEE cP=EEEE
D/CM2=EEEE
1/SEC=12.3 T=20.1C Z00:30
Under-Range Operation (<10% torque) (see Fig. 16):
? RPM=50 M=RV S=29 %=5.2 cP=1040
D/CM2=130
1/SEC=12.3 T=20.1C Z00:30
Negative Torque Operation (see Fig. 17):
RPM=50
M=RV S=29 %=-0.1 cP=---- D/CM2=----
1/SEC=12.3 T=20.1C Z00:30
M = Torque Range
T = Temperature
Z = Time
Figure II-18