Brookfield Engineering Labs., Inc.

Page 81

Manual No. M03-165-F0612

Calibration Procedure for a Small Sample Adapter

Brookfield recommends a two step check. First verify the calibration of the viscometer using the

standard viscometer spindles (LV #1-3, RV #2-6, HA #2-6 and HB #2-6 or cone/plate spindles) as

detailed in this appendix. Second verify the calibration of the viscometer using the Small Sample

Adapter. The use of an accessory device may increase the accuracy of measurement associated

with the DV-II+Pro.

When a Small Sample Adapter is used, the water jacket is connected to the water bath and the

water is stabilized at the proper temperature:

1) Put the proper amount of viscosity standard fluid into the sample chamber. The amount varies

with each spindle/chamber combination. (Refer to the Small Sample Adapter instruction

manual).

2) Place the sample chamber into the water jacket.
3) Put the spindle into the test fluid and attach the extension link, coupling nut and free hanging

spindle (or directly attach the solid shaft spindle) to the DV-II+Pro.

4) Allow 30 minutes for the viscosity standard, sample chamber and spindle to reach test temperature.
5) Measure the viscosity and record the Viscometer reading.

Note: The spindle must rotate at

least five (5) times before readings are taken.

Calibration Procedure for a Thermosel System

Brookfield recommends a two step check. First verify the calibration of the viscometer using the

standard viscometer spindles (LV #1-3, RV #2-6, HA #2-6 and HB #2-6 or cone/plate spindles) as

detailed in this appendix. Second verify the calibration of the viscometer using the Thermosel.

The use of an accessory device may increase the accuracy of measurement associated with the

DV-II+Pro.

When a Thermosel System is used, the controller stabilizes the Thermo Container at the test

temperature.

1) Put the proper amount of HT viscosity standard fluid into the HT-2 sample chamber. The

amount varies with the spindle used. (Refer to the Thermosel instruction manual).

2) Place the sample chamber into the Thermo Container.
3) Put the spindle into the test fluid and attach the extension link, coupling nut and free hanging

spindle (or directly attach the solid shaft spindle) to the DV-II+Pro.

4) Allow 30 minutes for the viscosity standard, sample chamber and spindle to reach test temperature.
5) Measure the viscosity and record the Viscometer reading.

Note: The spindle must rotate at

least five (5) times before readings are taken.

Calibration Procedure using UL or DIN Adapters

Brookfield recommends a two step check. First verify the calibration of the viscometer using the

standard viscometer spindles (LV #1-3, RV #2-6, HA #2-6 and HB #2-6 or cone/plate spindles)

as detailed in this appendix. Second verify the calibration of the viscometer using the UL or DIN

Adapters. The use of an accessory device may increase the accuracy of measurement associated

with the DV-II+Pro.