where the operator is instructed to mount a spindle 31 and then press a key to continue. Or you
might see:
SET TEMP TO 100C
PRESS A KEY
Figure IV-42
followed by:
WAIT TEMP 100C
P1 S00/12 00:14
Figure IV-43
Here, the B.E.V.I.S. program is waiting for the temperature to reach one hundred (100) °C before
it continues to the next program step. Also displayed is the program number (P1), the step
number and the total program steps (S00/12) and the elapsed time since the program (or step)
began (00:14). While in this “wait state” you can press the
OPTIONS/TAB
key to see the viscometer
default screen in Figure IV-44; the PGM will be flashing.
CP 123.4 55.1C
10 RPM PGM% 63.7
Figure IV-44
The DVLOADER Software is used to create, save, print, and down load programs to the DV-
II+Pro Viscometer. The next section explains how to use the DVLOADER software.