Manual No. M03-165-F0612
cP 123.5e6
MIN: 15 SEC: 13
Figure IV-20
Note:
When this mode has begun, a press of the MOTOR ON/OFF/ESCAPE key will
cancel the Timed Stop sequence and return the user to the screen of Figure IV-14.
Also note that data will be displayed in the currently selected method i.e. CGS
or SI units. Pressing the SELECT DISPLAY key allows display of alternate
data values such as Shear Stress, Shear Rate or Torque.
The seconds display will decrement from fifty-nine (59) to zero (0) in one (1) second intervals.
When seconds reaches zero (0), the minutes value will decrement by one (1) minute. This will
continue until all of the time has elapsed at which point the viscometer will display the following
screen:
cP 123e6
TIMED STOP DONE
Figure IV-21
At this point the viscometer will stop the motor and continue to display this screen until any key,
except the
UP
or
DOWN ARROW
key, the
key or the
SELECT DISPLAY
key, is pressed. The user
can, while this display is current, press the
UP
or
DOWN ARROW
keys to view the torque and speed
that were current at the time the display was frozen. The display would appear as follows:
%=76.4 RPM=100
TIMED STOP DONE
Figure IV-22
The display will switch between that of Figures IV-21 and IV-22 for each press of the
UP
or
DOWN
arrow keys. A press of the
screen would send one standard print line to the attached printer
for each press of the
key. Pressing any key (except the
UP
or
DOWN ARROW
keys, the
key or the
SELECT DISPLAY
key) will cause the viscometer to return to the Timed Stop Start Screen
of Figure IV-19 displayed awaiting another Timed Stop run.
Note: To run multiple timed stop tests, the user needs to press ENTER twice from the screen
shown as Figure IV-22.
The user can press the
key while in either of these two screens (Figures IV-21 and IV-22) to
send one standard print string to the attached printer as many times as the user presses the
key.
In addition, the
key can be pressed during the actual measurement to obtain instantaneous
data. Pressing any other key will exit this mode and return the viscometer to normal operation.