beautypg.com

Brookfield DV2+Pro Viscometer User Manual

Page 46

background image

Brookfield Engineering Labs., Inc.

Page 46

Manual No. M03-165-F0612

Click the Delete button to delete the B.E.V.I.S. command in the selected row of the

program grid.

Click the Up button to move the B.E.V.I.S. command in the selected row of the program

grid up one row.

Click the Down button to move the B.E.V.I.S. command in the selected row of the pro-

gram grid down one row.

Click the Clear button to clear the grid of all B.E.V.I.S. commands. Once cleared, the

commands cannot be retrieved.

Downloads a B.E.V.I.S. program from the PC to the DV-II+Pro

Exits the DVLOADER software program and returns the user to the WINDOWS Pro-

gram Manager.

V.4 Downloading a B.E.V.I.S. Program

Before down loading a B.E.V.I.S. program to the viscometer, ensure the following have been checked:

The appropriate cable (BEL Part# DVP-80) is connected between the selected COM port of

the PC and the viscometer.

The DV-II+Pro motor must be OFF.

Set PC PROG to “OFF.”

The Programmable DV-II+Pro is at the down load screen: OPTIONS|DOWNLOAD A

PROG|LOAD TO SLOT#x where x is slot 1,2,3, or 4. See Section III.6.

With the LOAD TO SLOT#x screen displayed, choose a storage slot using the DV-II+Pro

arrow keys then press the ENTER/AUTO RANGE key on the viscometer. If after 5 seconds,

the viscometer cannot communicate with the DVLoader program, the B.E.V.I.S. ERROR

NO PC ATTACHED message is displayed and a beeping sound is heard. If a connection is

established, the Down Load button on the PC software becomes enabled , and the DV-II+Pro

screen displays DOWNLOAD PROG TO STORAGE SLOT #1. Click on this button to

download the displayed program to the DV-II+Pro. When the down load is complete, the

DV-II+Pro displays DOWNLOAD DONE TO EXIT PRESS A KEY.

At this point, the program in the DV-II+Pro can be printed and/or run from the viscometer.

Click on this button to exit the DVLOADER software.

Note:

This is an important final step because it allows the RS232 port to

be used for output from the DV-II+Pro to a serial printer or PC.