Click the Delete button to delete the B.E.V.I.S. command in the selected row of the
program grid.
Click the Up button to move the B.E.V.I.S. command in the selected row of the program
grid up one row.
Click the Down button to move the B.E.V.I.S. command in the selected row of the pro-
gram grid down one row.
Click the Clear button to clear the grid of all B.E.V.I.S. commands. Once cleared, the
commands cannot be retrieved.
Downloads a B.E.V.I.S. program from the PC to the DV-II+Pro
Exits the DVLOADER software program and returns the user to the WINDOWS Pro-
gram Manager.
V.4 Downloading a B.E.V.I.S. Program
Before down loading a B.E.V.I.S. program to the viscometer, ensure the following have been checked:
•
The appropriate cable (BEL Part# DVP-80) is connected between the selected COM port of
the PC and the viscometer.
•
The DV-II+Pro motor must be OFF.
•
Set PC PROG to “OFF.”
•
The Programmable DV-II+Pro is at the down load screen: OPTIONS|DOWNLOAD A
PROG|LOAD TO SLOT#x where x is slot 1,2,3, or 4. See Section III.6.
With the LOAD TO SLOT#x screen displayed, choose a storage slot using the DV-II+Pro
arrow keys then press the ENTER/AUTO RANGE key on the viscometer. If after 5 seconds,
the viscometer cannot communicate with the DVLoader program, the B.E.V.I.S. ERROR
NO PC ATTACHED message is displayed and a beeping sound is heard. If a connection is
established, the Down Load button on the PC software becomes enabled , and the DV-II+Pro
screen displays DOWNLOAD PROG TO STORAGE SLOT #1. Click on this button to
download the displayed program to the DV-II+Pro. When the down load is complete, the
DV-II+Pro displays DOWNLOAD DONE TO EXIT PRESS A KEY.
At this point, the program in the DV-II+Pro can be printed and/or run from the viscometer.
Click on this button to exit the DVLOADER software.
Note:
This is an important final step because it allows the RS232 port to
be used for output from the DV-II+Pro to a serial printer or PC.