Brookfield DV2+Pro Viscometer User Manual
Page 38

Brookfield Engineering Labs., Inc.
Page 38
Manual No. M03-165-F0612
print mode and finally initiate either of the timed modes of operation. While setting
up the timed mode parameters, continuous print operation will cease. However, upon
starting the timed operation, the DV-II+Pro will output an initial data string to the printer
and then continue printing data strings (at the user defined time interval) for the duration
of the timed run. At the end of the timed run, continuous printing will again be disabled
and the user may print single strings (of the final data point) at your option until you exit
the timed mode. Upon returning to the default operation mode, continuous printing will
again resume at the user selected time interval. In a similar manner, if you are in the
once-per-
-key-press mode, when you enter the timed mode of operation you will
be able to print data strings at any time during the timed mode by pressing the
key.
IV.3.3 Print Time Interval
This option is used to set the print time interval to the selected printer. Scroll to Set Print Time,
as shown in Figure IV-29:
D TIMED TO TORQUE
SET PRINT TIME
Figure IV-29
Press the ENTER/AUTORANGE key. On entry, Figure IV-30 is displayed:
PRINT INTERVAL
SET MIN’S: 00
↑↓
Figure IV-30
Using the
UP
and
DOWN ARROW
keys, enter a value for the minutes between successive print strings.
This value can be as high as fifty-nine (59) minutes and as low as 00.
When satisfied, press the
OPTIONS/TAB
key to enter the seconds setting display:
PRINT INTERVAL
SET SEC’S: 00
↑↓
Figure IV-31
Using the
UP
and
DOWN ARROW
keys, enter a value for the seconds portion of the print interval.
This value can be between zero (0) and fifty-nine (59) seconds.
Note:
The value for minutes or seconds must be other than zero (0) or
you will print continuously when you exit this mode. A press of
the MOTOR ON/OFF/ESCAPE key would exit this option and
take you back to the screen of Figure IV-29.
Continued pressing of the
OPTIONS/TAB
key will toggle between the minutes and seconds input
screens. Press the
ENTER/AUTORANGE
key to accept the new values for print interval in minutes and
seconds. You will now be in the screen display of Figure IV-29 where you may re-enter the print
interval mode, or exit to the default screen (Figure II-5) by pressing the
MOTOR ON/OFF/ESCAPE
key.
#
$