Brookfield DV2+Pro Viscometer User Manual
Page 13
Brookfield Engineering Labs., Inc.
Page 13
Manual No. M03-165-F0612
II. GETTING STARTED
II.1 Autozero
Before readings may be taken, the Viscometer must be Autozeroed. This action is performed each
time the power switch is turned on. (Note: If cable DVP-80 is connected for printer or computer
communication see section II.9). The display window on the Viscometer will guide you through
the procedure as follows:
Turn the power switch (located on the rear panel) to the ON position. This will result in the screen
display shown in Figure II-1 indicating that the DV-II+Pro viscometer is in the standalone mode
(is not connected to a computer).
BROOKFIELD
DV-2+
PRO
VISCOMETER
Figure II-1
After a few seconds, the following screen appears indicating the version of the operating firmware
(the built in program which controls the instrument) and an alphanumeric code, which indicates
the Model number (see Table D-2 in Appendix D; the code indicates the spring torque rating or
the viscosity measurement range of your viscometer). For most DV-II+Pro Viscometers, this
information will be either “LV”, “RV” or “HB”:
BROOKFIELD
DV-2+
RV
V6.3
Figure II-2
No key press is required at this point. After a short time, the display will clear and the following
will be displayed:
REMOVE SPINDLE
PRESS ANY KEY
Figure II-3
After removing the spindle and pressing any key, the DV-II+Pro begins its Autozero. The screen
will flash “Autozeroing.”
After approximately 15 seconds, the display shows the screen in Figure II-4:
REPLACE SPINDLE
PRESS ANY KEY
Figure II-4
Pressing any key at this point results in the display of the DV-II
+
Pro default screen:
CP 0.0
20.1C
OFFRPM
% 0.0
Figure II-5