beautypg.com

Brookfield DV2+Pro Viscometer User Manual

Page 78

background image

Brookfield Engineering Labs., Inc.

Page 78

Manual No. M03-165-F0612

Table D-2 lists the model codes and spring torque constants for each Viscometer model.

Table D-2

MODEL

TK

MODEL CODE ON DV-II+

SCREEN

LVDV-II+

0.09373

LV

2.5LVDV-II+

0.2343

2.5 LV

5LVDV-II+

0.4686

5 LV

1/4 RVDV-II+

0.25

0.25 RV

1/2 RVDV-II+

0.5

0.5 RV

RVDV-II+

1

RV

HADV-II+

2

HA

2HADV-II+

4

2 HA

2.5HADV-II+

5

2.5 HA

HBDV-II+

8

HB

2HBDV-II+

16

2 HB

2.5HBDV-II+

20

2.5 HB

The full scale viscosity range for any DV-II+Pro model and spindle may be calculated using the

equation:

Full Scale Viscosity Range [cP] = TK * SMC * 10,000

RPM

where:

TK = DV-II+Pro Torque Constant from Table D-2

SMC = Spindle Multiplier Constant from Table D-1

The Shear Rate calculation is:

(Shear Stress (D/cm

2

)) = Viscosity (P) * Shear Rate (1/sec)

= TK * SMC * SRC * TORQ