Table D-2 lists the model codes and spring torque constants for each Viscometer model.
Table D-2
MODEL
TK
MODEL CODE ON DV-II+
SCREEN
LVDV-II+
0.09373
LV
2.5LVDV-II+
0.2343
2.5 LV
5LVDV-II+
0.4686
5 LV
1/4 RVDV-II+
0.25
0.25 RV
1/2 RVDV-II+
0.5
0.5 RV
RVDV-II+
1
RV
HADV-II+
2
HA
2HADV-II+
4
2 HA
2.5HADV-II+
5
2.5 HA
HBDV-II+
8
HB
2HBDV-II+
16
2 HB
2.5HBDV-II+
20
2.5 HB
The full scale viscosity range for any DV-II+Pro model and spindle may be calculated using the
equation:
Full Scale Viscosity Range [cP] = TK * SMC * 10,000
RPM
where:
TK = DV-II+Pro Torque Constant from Table D-2
SMC = Spindle Multiplier Constant from Table D-1
The Shear Rate calculation is:
(Shear Stress (D/cm
2
)) = Viscosity (P) * Shear Rate (1/sec)
= TK * SMC * SRC * TORQ