Brookfield DV2+Pro Viscometer User Manual

Page 9

background image

Brookfield Engineering Labs., Inc.

Page 9

Manual No. M03-165-F0612

Temperature Accuracy:

±1°C | -100°C to +149°C

±2°C | +150°C to +300°C

Operating Environment:

0°C to 40°C temperature range (32°F to 104°F)

20% - 80%R.H.: non-condensing atmosphere

Ball Bearing Option:

If you ordered the ball bearing suspension system with your new instrument please note

the following:

1) The ball bearing suspension in your Brookfield instrument is noted on the serial tag

on the back of the head by the letter “B” after the model.

2) When attaching and detaching the spindle, it is not necessary to lift the coupling where

the spindle connects to the instrument.

3) The Oscillation Check explained in Appendix K, Fault Diagnosis and Troubleshooting,

does not pertain to this instrument.

Electrical Certifications:

Conforms to CE Standards:

BSEN 61326:

Electrical equipment for measurement, control and laboratory use - EMC

requirements

BSEN 61010-1:

Safety requirements for electrical equipment, for measurement, control

and laboratory use

Notice to customers:

This symbol indicates that this product is to be recycled at an appropriate collection center.

Users within the European Union:

Please contact your dealer or the local authorities in charge of waste management on how to

dispose of this product properly. All Brookfield offices and our network of representatives and

dealers can be found on our website: www.brookfieldengineering.com

Users outside of the European Union:

Please dispose of this product according to your local laws.

I.4 Installation

Note: “IQ, OQ, PQ”, a guideline document for installation, operation and performance

validation for your DV-II+Pro digital viscometer can be downloaded from our

web site www.brookfieldengineering.com.

1) Assemble the Model S Laboratory Stand (refer to assembly instructions in Appendix I).

2) Put the viscometer on the stand.