beautypg.com

Brookfield DV2+Pro Viscometer User Manual

Page 39

background image

Brookfield Engineering Labs., Inc.

Page 39

Manual No. M03-165-F0612

Activating print selections in the Print mode can only be done by exiting to the main menu and

pressing the

PRINT

key for four (4) seconds. “P%” will flash in front of the torque reading, confirming

that you are now in the Print Interval mode. Pressing

PRINT

for one (1) second thereafter will

disable the Print mode and remove the “P%” from the display.

IV.3.4 PC Program (On/Off)

Note: This option does not apply to use with RHEOCALC software

.

This option causes the serial port of the DV-II+Pro viscometer to go into a high speed output

mode (approximately 3 print lines per second) for use with Brookfield WINGATHER® software

program. When ON, you may enter the Options menu but will not be allowed to make any option

selections until the PC PROG is turned OFF. All front panel keys will function normally when

you turn the option ON and return to normal viscometer operation by pressing the

MOTOR ON/

OFF/ESCAPE

key. When OFF, the DV-II+Pro will return to the last set print time interval when

printing is resumed.

Note:

WINGATHER® software can be used for data acquisition and

analysis. Rheocalc software can be used for complete external

control of DV-II+Pro and data acquisition.

From the Options menu, scroll to the screen shown in Figure IV-32:

SET PRINT TIME
D PC PROG OFF

Figure IV-32

Press the OPTIONS/TAB key to display Figure IV-33:

SET PRINT TIME
D PC PROG ON

Figure IV-33

Pressing the

OPTIONS/TAB

key would return you to the screen display of Figure IV-32. Repeated

pressing of the

OPTIONS/TAB

key would cause you to toggle back-and-forth between the displays

of Figure IV-32 and Figure IV-33.

To turn high speed output ON, press the

ENTER/AUTORANGE

key when the appropriate screen is

displayed. Then press the

MOTOR ON/OFF/ESCAPE

key to exit the Setup mode. This returns you

to the default screen display and resumes normal viscometer operation with high speed output

enabled and normal printer operation using the last entered print time interval.

Note:

For access to B.E.V.I.S. option, PC PROG must be OFF.

#
$

#
$