Brookfield DV2+Pro Viscometer User Manual
Page 48
Brookfield Engineering Labs., Inc.
Page 48
Manual No. M03-165-F0612
Command
Command Description
Parameter
Comments
WPT
Wait for % torque
85.0
wait until % torque = 85; a curing
cycle
Program 4: Spring relax
Command
Command Description
Parameter
Comments
WKY
Wait for a key press
Wind to 100%
tell operator to wind spindle
until 100% torque is reached
WPT
Wait for % torque
100.0
wait until 100% torque is reached
WKY
Wait for a keypress
Release spindle
tell operator to release the spindle
SPI
Set print interval
00:01
begin printing data at 1 second
intervals
WPT
Wait for % torque
0.0
wait for spindle to completely
unwind to 0% torque
Program 5: Variable speed
Command
Command Description
Parameter
Comments
SSN
Set viscometer speed
5.0
run at 50 RPM
WTI
Wait for time interval
00:10
wait for 10 seconds
PDN
Print data point now
print one data point
SSN
Set viscometer speed
10
run at 10 RPM
WTI
Wait for time interval
00:10
wait for 10 seconds
PDN
Print data point now
print one data point
SSN
Set viscometer speed
20
run at 20 RPM
WTI
Wait for time interval
00:10
wait for 10 seconds
PDN
Print data point now
print one data point
SSN
Set viscometer speed
5.0
run at 50 RPM
WTI
Wait for time interval
00:10
wait for 10 seconds
PDN
Print data point now
print one data point