The Configuration menu provides a means to setup up how the CO

2

Monitor functions. The

configuration items are saved and restored at start up of the CO

2

Monitor application. The

Configuration menu contains the following items:

Configure\ Mode\ Mainstream (Duet module only) - Checking this enables mainstream
operation and disables sidestream operation.

Configure\ Mode\ Sidestream (Duet module only) - Checking this enables sidestream
operation and disables mainstream operation.

Configure\ Display\Units - The CO

2

Monitor will display CO

2

values in either partial pressure

(mmHg) or percent. These units apply to the waveform and the EtCO

2

and InsCO

2

values.

Configure\ Display\Sweep Rate - The sweep rate setting determines the sweep rate of the CO

2

waveform. This actual sweep rate may vary depending on the PC’s configured display resolution.

Configure

Mode

Mainstream

Sidestream

Display

CO2 Module

Units

Sweep Rates

CO2 Scale

Gridlines

Waveform Fill

mmHg

Percent

6.25 mm/sec
12.5 mm/sec
25 mm/sec

50 mmHg

75 mmHg

100 mmHg

14%

10%

6%

Flow Rate

CommPort

90 ml/min

175 ml/min

150 ml/min