Two-point user calibration procedure – Welch Allyn LC101 CO2 Module - User Manual User Manual

Page 22

background image

LC101 CO

2

Module OEM Implementation Manual

Page 22

Confidential

Welch Allyn OEM Technologies

Two-Point User Calibration Procedure

Two-Point User Calibration Equipment Setup

Equipment needed:

calibration gas canister (8 – 12% CO

2

, ±0.02%, balance air or nitrogen)

gas valve and non-silicon tubing

sample elbow or tee

flow meter

sample line

watertrap

CO

2

scrubber (or 0% CO

2

medical-grade gas source)

CAUTION

Make sure that the tubing used to connect the gas supply to the LC101 does
NOT contain silicon. Silicon tubing absorbs CO

2

and affects the stability of

the gas concentration, thereby reducing the accuracy of the procedure.

Illustration of gas canister,
regulator and pneumatic circuit

10%
CO2

Bal. N2

Regulator

On/Off

Tee

Free Flow Away
From Monitor Input

To Monitor

Water Trap