LC101 CO

2

Module OEM Implementation Manual

Section 4 - Host/Module Communications

Welch Allyn OEM Technologies

Confidential

Page 35

Status Response Message Byte

<Sxxyy>

where yy

=

Status message

(2 Hex digits = 2 ASCII characters)

Status Response message byte (yy) contains:

generic status messages (00h – 06h)

fault status messages (10h – 85h)

Refer to the Status Message Table at the end of this section for message descriptions. More
detailed information about fault messages is also available in Appendix C, Error Messages &
Recovery.

Mode Command Responses

Host Command

Module Response

Module Response Format

Enter standby mode

xx = mode

yy = message
Example:

61= Standby mode

06= Acknowledge mode command

Enter measurement mode

xx = mode

yy = message
Example:

63= Measurement mode

06= Acknowledge mode command

Enter auto run mode

xx = mode

yy = message
Example:

64= Autorun mode

06= Acknowledge mode command