LC101 CO

2

Module OEM Implementation Manual

Page 40

Confidential

Welch Allyn OEM Technologies

Calibration Command Responses

Refer to the Status Message Code Table and the Host Command/Module Response Overview
Table at the end of this section for message descriptions.

Host Command

Module Response

Module Response Format

Update calibration date

xx = mode

yy = message
Example:

64= Autorun mode

24= Cal. ready for next step
Example:

64= Autorun mode

21= Cal. already in progress

Update span gas concen-
tration

xx = mode

yy = message
Example:

64= Autorun mode

24= Cal. ready for next step
Example:

64= Autorun mode

22= Cal. not in progress

Sample zero reference gas

xx = mode

yy = message
Example:

64= Autorun mode

25= Cal. in progress

Sample span gas

xx = mode

yy = message
Example:

64= Autorun mode

25= Cal. in progress

Abort calibration

xx = mode

yy = message
Example:

64= Autorun mode

00= OK (operation aborted)