Unexpected forward flow, Internal disconnects, Watertrap receptacle – Welch Allyn LC101 CO2 Module - User Manual User Manual

Page 50: Sensor inlet/outlet, Flow control circuitry, Pump inlet

background image

LC101 CO

2

Module OEM Implementation Manual

Page 50

Confidential

Welch Allyn OEM Technologies

Unexpected Forward Flow

This condition occurs when the pneumatic input is connected to a positive pressure source, or
flow source. The flow control circuitry checks the Pneumatic System for this condition prior to
turning ON the pump. If a positive pressure exists on the exhaust prior to the pump starting, the
Module reverts to the fault mode. During this condition, the pump is not operational and the
Module responds to status requests with Unexpected Forward Flow.

Internal Disconnects

Internal disconnects which may occur include

watertrap receptacle

sensor inlet/outlet

flow control circuitry

pump inlet

pump exhaust

Watertrap Receptacle

If a pneumatic connection at the watertrap receptacle is loose or has failed, the transmission of
CO

2

waveform and breath data packets is halted. The Module does not report this fault to the

host but clears the breath data based on the default specified in the No Breath Timeout
command.

Sensor Inlet/Outlet

If a pneumatic connection at the sensor inlet or outlet is loose or has failed, the transmission of
CO

2

waveform and breath data packets is halted. The Module does not report this fault to the

host but clears the breath data based on the default specified in the No Breath Timeout
command.

Flow Control Circuitry

If a pneumatic connection in the flow control circuitry is loose or has failed, transmission of CO

2

waveform and breath data packets is halted. Pump operation may be unstable due to the lack of
flow control feedback. The Module reverts to fault mode and may respond to status requests by
issuing a number of different fault codes.

Pump Inlet

If the pneumatic connection at the pump inlet is loose or has failed, transmission of CO

2

waveform and breath data packets is halted. The set flow rate cannot be achieved since the
pump is disconnected. The Module reverts to the fault mode and responds to status requests
with an Exhaust Occlusion response.