Zero calibration sequence example – Welch Allyn LC101 CO2 Module - User Manual User Manual

Page 57

background image

LC101 CO

2

Module OEM Implementation Manual

Appendix A - Software Prodedures

Welch Allyn OEM Technologies

Confidential

Page 57

1.

Initialize breath data update rate .

2.

Request ambient barometric pressure .

3.

Divide ETCO

2

(in mmHg) value by ambient barometric pressure (in mmHg). The result is the

ETCO

2

value as a percentage of the barometric pressure.

Zero Calibration Sequence Example

Seq.

Host

Module

Description

1

Host sends Unprotect command.

Module responds to command from host.

2

Host sends

Perform Zero Calibration command.

Module responds with status -

Calibration in Progress, or

if not in unprotected measurement or autorun mode, response is

Invalid

Operation (use ), or
if the previous calibration is active, response is

Cal. Already in Progress,

or

if the LC101 had been operating for < 5 min., response is

Cal. Low Run

Time.

3

Host polls for completion of sampling.

Module responds with status –

Calibration in Progress. Repeat this

step until a response other than is returned.

Module responds with status –

Calibration OK, or

if acquiring data does not meet specifications, response is

Unable to

Perform Cal., Data Error.

4

Host sends Protect command.

Module responds to command from host.