beautypg.com

Welch Allyn LC101 CO2 Module - User Manual User Manual

Page 37

background image

LC101 CO

2

Module OEM Implementation Manual

Section 4 - Host/Module Communications

Welch Allyn OEM Technologies

Confidential

Page 37

Request single breath packet

xx = ETCO

2

mmHg

yy = RR bpm

ww = InsCO

2

mmHg

Example:

ETCO

2

= 32

RR= 21

InsCO2= 0

Request current pump flow rate

xxxx = pump flow rate ml/min

Example:

flow rate = 150 ml/min

Request sensor EEPROM revi-
sion #

xx = revision #

Example:

revision # = 2

Request sensor customer code

xx = customer code

Example:

customer code = 00

Request sensor serial #

xxxx = serial #

Example:

serial # = 13506

Request last calibration date

mm = month

dd = day

yyyy = year

Example:

last cal date= 6/4/1998