Module OEM Implementation Manual

Page 72

Confidential

Welch Allyn OEM Technologies

System Faults

The majority of system faults are conditions that are internal to the Module and not caused by the
host. The exceptions are 81h, 82h, 84h and 85h.

To clear most system faults:

Do a reset (with the exception of 81, 82, 84 and 85).

System faults are listed in the following table:

Status Message
Code (hex)

Definition

Possible Causes

40

watchdog error

• watchdog timeout

44

system EEPROM error
in CCITT/CRC code

• invalid system EEPROM CCITT/CRC code

46

system FLASH ROM
error in CCITT/CRC
code

• invalid FLASH ROM CCITT/CRC code

• faulty FLASH ROM chip

47

system communication
error

• microprocessor failure to complete task

• malfunctioning real-time operating system

4B

system external RAM
error

• faulty system external RAM

4C

system internal RAM
error

• faulty system internal RAM

4D

system FLASH ROM
error in checksum

• invalid system internal FLASH data

• faulty FLASH ROM chip

4E

stack overflow error

• stack usage error

4F

system main program
exited

• executive failure

80

pump failure

• faulty pump;

• pneumatics disconnected

81

unexpected reverse flow

• watertrap/cannula/exhaust tubing not properly con-

nected

82

unexpected forward flow

• watertrap/cannula/exhaust tubing not properly con-

nected

84

barometric pressure too
high

• host system operating above 824 mmHg (1,150 ft

below sea level)

• faulty barometric transducer

85

barometric pressure too
low

• host system operating below 408 mmHg (15,000 ft

above sea level)

• faulty barometric transducer