System faults – Welch Allyn LC101 CO2 Module - User Manual User Manual
Page 72

LC101 CO
2
Module OEM Implementation Manual
Page 72
Confidential
Welch Allyn OEM Technologies
System Faults
The majority of system faults are conditions that are internal to the Module and not caused by the
host. The exceptions are 81h, 82h, 84h and 85h.
To clear most system faults:
•
Do a reset (with the exception of 81, 82, 84 and 85).
System faults are listed in the following table:
Status Message
Code (hex)
Definition
Possible Causes
40
watchdog error
• watchdog timeout
44
system EEPROM error
in CCITT/CRC code
• invalid system EEPROM CCITT/CRC code
46
system FLASH ROM
error in CCITT/CRC
code
• invalid FLASH ROM CCITT/CRC code
• faulty FLASH ROM chip
47
system communication
error
• microprocessor failure to complete task
• malfunctioning real-time operating system
4B
system external RAM
error
• faulty system external RAM
4C
system internal RAM
error
• faulty system internal RAM
4D
system FLASH ROM
error in checksum
• invalid system internal FLASH data
• faulty FLASH ROM chip
4E
stack overflow error
• stack usage error
4F
system main program
exited
• executive failure
80
pump failure
• faulty pump;
• pneumatics disconnected
81
unexpected reverse flow
• watertrap/cannula/exhaust tubing not properly con-
nected
82
unexpected forward flow
• watertrap/cannula/exhaust tubing not properly con-
nected
84
barometric pressure too
high
• host system operating above 824 mmHg (1,150 ft
below sea level)
• faulty barometric transducer
85
barometric pressure too
low
• host system operating below 408 mmHg (15,000 ft
above sea level)
• faulty barometric transducer