Lc101 module interface connector, Watertrap switch connection, Pump – Welch Allyn LC101 CO2 Module - User Manual User Manual
Page 16: Watertrap receiver

LC101 CO
2
Module OEM Implementation Manual
Page 16
Confidential
Welch Allyn OEM Technologies
LC101 Module Interface Connector
The LC101 Module interface connector is a 4-pin connector (Molex P/N 22-11-2042). The host
may use a 4-pin polarized housing (Molex P/N 22-01-3047 with Molex 2759 series terminals) or
equivalent.
Watertrap Switch Connection
Watertrap switch connection is via a 2-pin Molex connector on the Main Board.
Pump
The pump is mounted on the Main Board. The pump wiring harness is connected to the Main
Board via a 2-pin connector (Molex P/N 22-01-3027).
The pump wiring meets typical insulation standards but does not meet patient isolation
requirements. If the pump must be moved to a different location, patient isolation requirements
must not be compromised. Refer to Section 6, Regulatory, for additional regulatory and safety
information.
Watertrap Receiver
The internal space available in the host monitor must be considered when incorporating the
watertrap receiver into the host system design. To accommodate various OEM dimensions, two
watertrap receivers are offered: long and short. The long watertrap receiver requires more
intrusion into the host monitor but the watertrap does not protrude as far externally from the host
chassis than with the short watertrap receiver.
Pin
Signal
Signal Description
1
Vin
a
a.
The “square” plated through-hole designates pin 1 on the Main Board.
input power
2
gnd
ground
3
TXD
data from Main Board to host system
4
RXD
data from host system to Main Board