beautypg.com

Module responses – Welch Allyn LC101 CO2 Module - User Manual User Manual

Page 33

background image

LC101 CO

2

Module OEM Implementation Manual

Section 4 - Host/Module Communications

Welch Allyn OEM Technologies

Confidential

Page 33

Zero Calibration

The Zero Calibration adjusts the IR source current to compensate for detector output changes.
The host is required to inform the LC101 Module when to start sampling 0% CO

2

gas.

When performing a Zero Calibration, the LC101 Module must first be in either Unprotected
'Measurement' or 'Autorun' mode. During the calibration, CO

2

waveform and Breath data will still

be sent to the host.

To get an accurate calibration, the LC101 Module needs to be operating for at least 5 minutes in
measurement or autorun modes. The LC101 will not allow a calibration if this requirement has
not been met. In addition, the calibration should be performed at room/ambient temperature.

After a successful Zero Calibration, the Module updates the zero residual offset constants
located in the

C-Cap EEPROM Memory Map.

Module Responses

The host must initiate all communication with the Module with the exception of periodic CO

2

waveform and breath data packets. The type of response sent is determined by the type of host
command. Responses include:

status responses

mode command responses

simple command responses

configuration command responses

calibration command responses

Host Command

Description

Perform Zero Calibration

Abort Calibration