beautypg.com

Correction for desflurane, Summary of o2, n2o, and desflurane compensation, Percent co2 calculation – Welch Allyn LC101 CO2 Module - User Manual User Manual

Page 56: Percent co, Calculation, Summary of o, O, and desflurane compensation

background image

LC101 CO

2

Module OEM Implementation Manual

Page 56

Confidential

Welch Allyn OEM Technologies

Correction For Desflurane

The effect of desflurane on the CO

2

measurement is similar to the effect of N

2

O. For desflurane

concentrations above 12%, the N

2

O correction may be used.

Summary of O

2

, N

2

O, and Desflurane Compensation

The LC101 Module compensations for O

2

, N

2

O and desflurane are OFF by default and must be

enabled by the host via menu selection or other means. The variables and correction equations
are summarized in this table.

The O

2

/N

2

O/desflurane compensations are enabled/disabled by the following Function Enable/

Disable commands.

Percent CO

2

Calculation

It is necessary to convert the CO

2

display from mmHg to percent CO

2

in order to compare the

measurement to a calibrated gas, usually marked in percent CO

2

. (There are other user

environments where the preferred units of measurement is % CO

2

.) The host system is

responsible to convert partial pressure (mmHg) CO

2

to the corresponding percentage. Typically,

calibrated gas measurement is a steady state measurement, so updated breath data is not
necessary. Also, calibrated gases are dry gases; therefore the water vapor and BTPS
compensations should be disabled by the host in those instances where dry gas readings are
made.

O

2

Correction

N

2

O/Desflurane Correction

Operating Conditions

OFF

OFF

O

2

≤ 50%, no N

2

O

ON

OFF

O

2

> 50%, no N

2

O

OFF

ON

O

2

≤ 50% and N

2

O or desflurane

≥ 12%

ON

ON

O

2

> 50% and N

2

O or desflurane

≥ 12%

Compensation

Function Enable Command

O

2

compensation OFF

O

2

compensation ON

N

2

O compensation OFF

N

2

O compensation ON

Desflurane compensation OFF