beautypg.com

Zero calibration procedure – Welch Allyn LC101 CO2 Module - User Manual User Manual

Page 20

background image

LC101 CO

2

Module OEM Implementation Manual

Page 20

Confidential

Welch Allyn OEM Technologies

To get an accurate calibration, the LC101 needs to be operating for at least 5 minutes in
measurement or autorun mode with the scrubber installed. The LC101 internally monitors the
run time and will not allow a Zero Calibration if the run time requirement has not been met.

The LC101 monitors the calibration progress and provides a “Calibration OK” (or “not OK”)
message to the host system when polled for status. See the LC101 Software Protocol and the
Zero Calibration Commands and Responses for more detail. After a successful Zero Calibration,
the LC101 updates the zero residual offset constants located in the LC101 C-Cap EEPROM
Memory and the IR source current.

Zero Calibration Procedure

Equipment needed:

watertrap

CO

2

scrubber (or 0% CO

2

medical-grade gas source; see the Two-Point User

Calibration procedure for pressurized gas setup)

To Perform a Zero Calibration:

The user is required to apply a 0% CO

2

gas, or “zero” gas, during the steps of a Zero Calibration.

To perform Zero Calibration, the user must follow the Original Equipment Manufacturer’s
procedure typically provided in the Operator’s Guide or User Service Manual.

1.

Select a well ventilated room to perform the calibration.

2.

Make sure the LC101 has been operating for at least 5 minutes prior to the Zero Calibration.

3.

Attach the CO

2

scrubber to the watertrap inlet according to the CO

2

scrubber “Directions for

Use” (or attach a 0% medical-grade gas source). Now let the LC101 operate for one
minute.

4.

After approximately one minute, observe the CO

2

reading. The CO

2

reading should be

between 0.0% - 0.3% with well-ventilated room air.

5.

Proceed with the zero gas phase of the calibration as defined by the host system.

6.

Disconnect the CO

2

scrubber (or gas source) from the watertrap after the zero gas calibra-

tion.