Example of Verification Using Dry Gas

Because Module verification is typically accomplished using “dry” CO

2

gas from a canister, the

automatic compensations for water vapor and BTPS must be disabled prior to verification. The
host system can temporarily disable these module functions by using the sequence below:

NOTE

The Module will revert to the default settings (BTPS and Water Vapor compensations
enabled) when the Module is reset or power cycled. Refer to the Host/Module
Communications section (page 25
) for more details on the commands listed in the
table. Because the CO

2

concentration contained in a canister or cylinder is typically

listed in percent values, it is recommended the host also convert the module’s CO

2

data

to percent for comparison purposes.

6

Host sends

Sample CO2 Span Gas command.

Module responds with status -

Calibration in Progress, or

if not in unprotected measurement or autorun mode, response is

Invalid

Operation (use ), or
if a calibration had not been started, response is

Cal. Not in Progress, or

if the calibration has exceeded its maximum time allowed, response is
Unable to Perform Cal., Missing Parameters.

7

Host polls for completion of sampling.

Module responds with status -

Calibration in Progress. Repeat this

step until a response other than is returned.

Module responds with status -

Calibration OK, or

if the calibration has exceeded its maximum time allowed, response is
Unable to Perform Cal., Missing Parameters, or
if acquiring data does not meet specifications, response is

Unable to

Perform Cal., Data Error.

8

Host sends Protect command.

Module responds to command from host

Command from Host

Description

Note

C00

Requests Module status

Verify no faults are reported

Q7F0

Unprotects the Module

Q080

Disables Water Vapor Comp

Q090

Disables BTPS Compensation

O00

Disable Breath Data Packet

For continuous CO2 mode