beautypg.com

Appendix e specifications, Lc101 main board, Appendix e – Welch Allyn LC101 CO2 Module - User Manual User Manual

Page 77: Specifications

background image

LC101 CO

2

Module OEM Implementation Manual

Welch Allyn OEM Technologies

Confidential

Page 77

Appendix E

Specifications

LC101 Main Board

PHYSICAL SPECIFICATIONS

Size

3.15 in. x 3.90 in. (80.3 mm x 99.3 mm)

Clearance

0.85 in. (22 mm) above board, 0.12 in (3 mm) below board

Orientation

No limitations

Weight

Less than 1 pound (454 grams) with full pneumatics.

SYSTEM PERFORMANCE SPECIFICATIONS

Sampling Method

Built-in sidestream sensor

Patient Application

Neonatal to adult

CO

2

Concentration Display Range

0 - 99 mmHg

Respiration Rate Range

1 - 99 breaths per minute

Typical CO

2

Accuracy

a

Start-Up Time

Less than 10 seconds to acquire CO

2

waveform data

Less than 3 minutes to full operating specification.

Calibration

Zero Calibration and Two-Point User Calibration

C-Cap Rise Time

Less than 200 msec (10% - 90%, bench only)

Flow Rate Range

User-selectable, variable from 90 to 200 ml/min (defaults to 175
ml/min)

Flow Rate Accuracy

-20/+15% of set value.