Module OEM Implementation Manual

Page 38

Confidential

Welch Allyn OEM Technologies

Configuration Command Responses

Configuration responses are data packets sent by the Module to the host at programmed
intervals. A single host measurement or auto run mode command initiates unsolicited
transmission of CO

2

waveform and breath data packets. The programmed interval between

packets is either defined by the host via a configuration command or the host may use the default
setting.

, CO

2

Waveform Data

Includes 4 bytes of CO

2

waveform data that are sent at the programmed interval during

measurement and auto run modes.

where xxyy:

xxyy

CO

2

waveform data mmHg

× 256

4*(ASCII)

range:

0 – 99.996 mmHg (0000h – 63FFh)

Example:

CO

2

= 37.50 mmHg

Module Response

Module Response Format

Unsolicited CO

2

waveform packet at programmed interval

see below

Unsolicited breath data packet at programmed interval

see below

{STX}

“W”

xx

yy

CCITT/CRC

{ETX}

waveform byte string

0x25

0x80

0x79

{STX}W258079{ETX}

02h

57h

32h35h

38h30h

37h39h

03h