beautypg.com

Pump exhaust, Pump failure, Low flow – Welch Allyn LC101 CO2 Module - User Manual User Manual

Page 51: High flow, No flow, Low flow high flow no flow

background image

LC101 CO

2

Module OEM Implementation Manual

Section 5 - Pneumatic Operation

Welch Allyn OEM Technologies

Confidential

Page 51

Pump Exhaust

If a pneumatic connection at the pump exhaust is loose at either the pump exhaust or internal
monitor exhaust port, gas build up within the host monitor can occur. The Module does not report
this fault to the host.

WARNING When using oxygen or flammable anesthetics, a pump exhaust leak can

present a fire hazard. It is the responsibility of the OEM to meet regulatory
requirements with regard to single fault protection and exhaust tubing.

Refer to Section 6, Regulatory, for additional regulatory and safety information.

Pump Failure

Events resulting from pump failure include:

low flow

high flow

no flow

Low Flow

The pump may fail to meet the set flow rate due to low flow caused by a nonfunctional pump. The
Module responds to status requests with a Exhaust Occlusion Or Pneumatic Leak
response.

NOTE

The Module remains indefinitely in the total exhaust occlusion state if not commanded
by the host.

High Flow

The pump may fail to meet the set flow rate due to high uncontrollable flow caused by a runaway
pump. The Module reverts to the fault mode, turns OFF the pump and responds to status
requests with an Pump Failure response.

No Flow

The pump may fail to meet the set flow rate due to no flow caused by a nonfunctional or
electrically disconnected pump. The Module responds to status requests with a
Exhaust Occlusion Or Pneumatic Leak response.

WARNING The Module remains indefinitely in the total exhaust occlusion state if not

commanded by the host.