beautypg.com

Section 2 lc101 module interface, Power requirements, Lc101 main board – Welch Allyn LC101 CO2 Module - User Manual User Manual

Page 15: Section 2, Lc101 module interface, Power requirements lc101 main board

background image

LC101 CO

2

Module OEM Implementation Manual

Welch Allyn OEM Technologies

Confidential

Page 15

Section 2

LC101 Module Interface

Power Requirements

LC101 Main Board

Description

Typical

Input voltage range

5 V regulated, or +5.75 to +14.50 Vdc

Ripple 100 mV peak-to-peak

Typical non-measurement mode power

570 mW at 8 Vdc

Typical operating power

1.3 W at 8 Vdc

Typical occluded power

1.5 W at 8 Vdc

Description

Specification

Dimensions

3.15 in. L x 3.90 in. W x 1.00 in. H

(80.3 mm

× 99.3 mm × 25mm)

Orientation

No limitations

Weight

< 1 lb (454 gm)