29, Unable to Perform Calibration, Data Error
The acquired CO
2
data is invalid or out of range for the calibration. The Module will abort the
calibration. This error also occurs when calibration is aborted due to any other non-fatal error.
2A, Bad Calibration CRC
The system EEPROM CRC is not valid after being updated with the new constants.
Sensor Faults
CO
2
sensor faults are conditions that result in shutting down the sensor. A CO
2
sensor shutdown
turns off the lamp and the pump.
To clear a sensor fault, perform a software reset followed by a new host command.
If the Module
does not accept the command, a hard reset is required.
Sensor faults are listed in the following table.
51, CO
2
Sensor Manufacturer Code Mismatch
Sensor manufacturer code stored in sensor EEPROM does not match the host monitor
manufacturer code.
•
Verify the data in the EEPROM location.
Status
Message Code
(hex)
Definition
Possible Causes
51
Customer code mismatch
• Incorrect EEPROM data or bad sensor EEPROM
57
CO
2
sensor not found
• Main Board not acknowledging the sensor
• loose or faulty sensor connection
60
CO
2
sensor EEPROM error
in revision #
• incompatible/invalid sensor EEPROM revision #
• faulty sensor EEPROM
65
CO
2
sensor EEPROM error
in read or write operation
• invalid read/write to sensor EEPROM;
• faulty sensor EEPROM
66
CO
2
sensor EEPROM error
in CCITT/CRC code
• invalid sensor EEPROM CCITT/CRC code;
• faulty sensor EEPROM
70
CO
2
sensor temperature
too high
• sensor exposure to excessive heat source;
• faulty sensor
71
CO
2
sensor temperature
too low
• sensor exposure to excessive cold source;
• faulty sensor