LC101 CO

2

Module OEM Implementation Manual

Page 70

Confidential

Welch Allyn OEM Technologies

29, Unable to Perform Calibration, Data Error

The acquired CO

2

data is invalid or out of range for the calibration. The Module will abort the

calibration. This error also occurs when calibration is aborted due to any other non-fatal error.

2A, Bad Calibration CRC

The system EEPROM CRC is not valid after being updated with the new constants.

Sensor Faults

CO

2

sensor faults are conditions that result in shutting down the sensor. A CO

2

sensor shutdown

turns off the lamp and the pump.

To clear a sensor fault, perform a software reset followed by a new host command.

If the Module

does not accept the command, a hard reset is required.

Sensor faults are listed in the following table.

51, CO

2

Sensor Manufacturer Code Mismatch

Sensor manufacturer code stored in sensor EEPROM does not match the host monitor
manufacturer code.

Verify the data in the EEPROM location.

Status
Message Code
(hex)

Definition

Possible Causes

51

Customer code mismatch

• Incorrect EEPROM data or bad sensor EEPROM

57

CO

2

sensor not found

• Main Board not acknowledging the sensor

• loose or faulty sensor connection

60

CO

2

sensor EEPROM error

in revision #

• incompatible/invalid sensor EEPROM revision #

• faulty sensor EEPROM

65

CO

2

sensor EEPROM error

in read or write operation

• invalid read/write to sensor EEPROM;

• faulty sensor EEPROM

66

CO

2

sensor EEPROM error

in CCITT/CRC code

• invalid sensor EEPROM CCITT/CRC code;

• faulty sensor EEPROM

70

CO

2

sensor temperature

too high

• sensor exposure to excessive heat source;

• faulty sensor

71

CO

2

sensor temperature

too low

• sensor exposure to excessive cold source;

• faulty sensor